Table of Contents

Chapter 1 Preface

1.1 Overview ············································ 3
1.2 Introduction ············································ 3
1.3 Environmental parameters ········································ 3
1.4 Power supply ············································ 3
1.5 Load cell ···································· 3
1.6 Mainboard digital input port ······························· 3
1.7 Mainboard digital output port ······························· 3
1.8 Communication board ········································ 3

Chapter 2 Installation
2.1 Overview ············································ 4
2.2 Installation ············································· 4
2.3 Safety Warning ········································· 4
2.4 Wiring ············································ 4
2.5 Initial settings of the instrument ······························· 5
2.5.1 Setting the cumulative unit ······························· 5
2.5.2 Select flow unit ······························· 6
2.5.3 Set the maximum scale flow ································ 6
2.5.4 Select scale division ···································· 6
2.5.5 Speed signal input form ····························· 6
2.5.6 Select calibration mode ······························· 7

2.5.7 Enter the calibration constant ······························· 7
2.5.8 Establish a test cycle ······························· 7
2.5.9 Automatic zero-point calibration ································ 9
2.5.10 Automatic interval calibration ······························· 10

Chapter 3 Operation Details
3.1 Front panel ········································· 11
3.1.1 LED status indication ····································· 11
3.1.2 Keyboard ··········································· 11
3.2 Menu display 12
3.3 Operation display 12
3.3.1 Main accumulation ········································· 13
3.3.2 Reset accumulation ····································· 13
3.3.3 Alarm processing ···································· 13
3.4 Calibration 15

Chapter 4 Menu
4.1 Main menu 1 15
4.1.1 Read current zero point and manual zero point ······················ 15
4.1.2 Automatic zero-point calibration ···············15
4.1.3 Reading interval and manual calibration interval ···············15
4.1.4 Automatically adjust the interval… 16
4.1.5 Physical calibration ······································ 16
4.2 Main Menu 2 ··········································· 17
4.2.1 Display ········································· 17
4.2.2 Select the accumulation unit ······························· 17
4.2.3 Select flow unit ································· 18
4.2.4 Set display flow buffer time ··························· 18
4.2.5 The third line display······························ 18
4.2.6 Scale data ········································· 18
4.2.7 Set the maximum scale capacity ······························· 19
4.2.8 Select scale division ······································· 19
4.2.9 Belt scale model······································· 19
4.2.10 Set zero dead zone ·································19
4.2.11 Remote cumulative output indexing ······························· 19
4.2.12 Cumulative pulse width····························20
4.2.13 Load cell capacity································20
4.2.14 Load cell sensitivity···························20
4.2.15 Number of load cells·····························20
4.2.16 Define the speed signal input mode ························· 21
4.2.17 Resistance value of load cell······························· 21
4.2.18 Idle distance ··································· 21
4.2.19 Max kg/m ···································· 21
4.2.20 Inclination angle of belt scale ···································22
4.2.21 Calibration data ······································· 22
4.2.22 Calibration mode ···············22
4.2.23 Choosing the weight to be calibrated by hanging code ···············22
4.2.24 Hanging Code Calibration Constant···································· 22
4.2.25 Automatically establish a test cycle ··············· 23
4.2.26 Automatic zero tracking ···············23
4.3 Main Menu 3 ·········································· 23
4.3.1 Diagnosis 23
4.3.2 Protection settings…24
4.3.3 Change Password·······························24
4.4 Main Menu 4 ·········································· 24
4.4.1 Input Setting ······································ 25
4.4.2 Output setting ······································ 25
4.4.3 Alarm setting ······································ 26
4.4.4 High flow setpoint ··································26
4.4.5 High flow alarm delay ·······························26
4.4.6 Low flow alarm point ·································· 27
4.4.7 Low flow alarm delay ······························ 27

4.4.8 Control deviation setpoint ······························ 27
4.4.9 Control deviation alarm delay ······························· 27
4.4.10 Definition of Alarm ····································· 28
4.4.11 Alarm reset selection····························· 28
4.5 Main Menu 5
4.5.1 Baud rate setting ···································· 29
4.5.2 Address Setting
4.5.3 Function selection ···································· 29
4.6 Main Menu 6 ·········································· 30
4.6.1 Setting the scale factor ··································· 30
4.6.2 Set integral time ···································30
4.6.3 Setting the differential time ···············30

Chapter 1 Preface

1.1Overview
This operation manual provides detailed information about the installation, operation, calibration, and maintenance of the belt scale conveyor controller.

1.2 Introduction
The feeder controller includes on-site installation and panel installation. It adopts microprocessor control to process the weight signal and belt speed signal of the load cell, calculate the flow rate and cumulative amount of the flowing material, and display it on the instrument. At the same time, the instrument has a built-in PID regulator, which provides a standard 4-20mA output to control the operation of the feeder. And provide remote pulse count output. An RS-232 or RS-485 interface can be added to communicate with the host computer using a communication board. The instrument adopts an all-Chinese operation interface, which is intuitive and easy to operate. And has a variety of automatic detection functions, which is convenient for users to calibrate and maintain.

1.3 Environmental parameters
1. Indoor/Outdoor The installation location should be close to the load cell. Pay attention to dust and moisture.
2. Storage temperature -40℃ — +70℃
Operating temperature -10℃ — +50℃
3. Maximum relative humidity: 95%

1.4 Power supply
·220VAC±10% 50Hz
·Fuse 2A
·EMI/RFI filter

1.5 Load cell
·The instrument provides 10VDC ± 10%, 200mA excitation power, 4 load cells can be connected in parallel
·Sensitivity, 0.5mV/V – 3.5mV/V
· Maximum input signal 33mV
· Weighing sensor shield grounding
· Use excitation compensation circuit when the cable distance is greater than 60 meters (not more than 900 meters)
Jumper selection local or remote compensation

1.6 Mainboard digital input port
The instrument mainboard provides 3 programmable input ports to receive dry contact switch signals.

1.7 Mainboard digital output port
The mainboard of the instrument provides 3 programmable output ports, 24VDC open collector output. It can directly drive the control relay.

1.8 Communication board
The instrument can be installed with a communication board and can use standard RS-232, RS-485 serial interface to communicate with the host computer to collect instrument data and operate the instrument.

Chapter 2 Installation

2.1 Overview
This chapter includes the installation, configuration, and initial programming of the feeder controller.
The instrument should not be installed in a place where the environmental conditions change drastically. Try to install it indoors, otherwise, it is necessary to install protective facilities to avoid direct sunlight, humidity, collisions, and severe mechanical vibrations. The meter can be installed at the farthest distance of 900 meters from the load cell.

2.2 Installation
The wall-mounted integrator is vertically installed on a solid, flat wall through the four mounting holes at the back of the chassis.

2.3 Safety warning
warn
Do not perform any installation, operation, and maintenance operations before reading the following warning regulations
1. Do not connect to the power supply of the meter until you fully understand the manual.
2. Strictly follow the warnings and operating procedures in this manual to avoid personal injury and equipment damage.

2.4 Wiring
1. Make sure the power is off
2. Do not route signal wires and power wires or other cables with strong electromagnetic interference in the same conduit.
3. The instrument housing and conduit should be grounded at one point.
4. Check that all connections are correct and secure.
5. It is strictly forbidden to use a “megohm meter” to check the circuit.
6. A circuit breaker (maximum 20A) should be installed in the power supply line. The installation position of the circuit breaker should be easy to operate and clearly marked for this instrument.
7. All ducts should enter from the bottom of the chassis, not from the top or side of the chassis.

conveyor belt scales

Notice:
1. Do not shorten the cable of the load cell
2. The sensor signal line should not use the same conduit as the power line.
3. The shielding layer of the shielded cable should be grounded at one end and connected as shown in the figure.
4. The chassis must be grounded.
2.5 Initial settings of the instrument
After the mechanical and electrical installations are completed, initialize the instrument and program it. The following parameters should be entered before calibration.
2.5.1 Set accumulation unit
Press the Menu key twice until the screen displays the following information

conveyor belt scales for sale

weigh shark conveyor belt scales

conveyor belt weighing systems

conveyor belt weighing systems

 

2.5.6 Automatic zero-point calibration
During automatic zero-point calibration, the belt scale runs at maximum speed
Press the menu button to display the main menu 1 on the screen

— Main menu 1 —
Press [Menu] to continue
Zero Interval Object
Calibration Calibration Calibration

Calibrate by zero point, the screen shows:

–Automatic zero-point calibration —
Test-run the belt conveyor without loading material
Press the start button again
Automatic Exit Manual

Press the automatic button, the screen shows:

–Automatic zero-point calibration —
Test-run the belt conveyor without loading material
Press the start button again
Start Exit Manual

Press the start button, the screen shows:

Automatic Zero Set
Time Remaining
XXX
XXX.X ton/hour
Grand Total XXXX ton

When the time remaining is 0, the screen shows:

Automatic Zero Set Done
Error XXX.X %
Change the zero point?
Change Exit

If you need to change the zero point, press Change, the screen shows:

2.5.10 Automatic interval calibration
After the zero-point calibration is completed, press the Menu key to return to the main menu 1, press interval calibration, the screen shows:

— XX Calibration interval — Start to run the belt
Press the start button again
Start Exit Manual

Interval calibration can use three different analog load calibration methods: electronic calibration, hanging code calibration, and chain code calibration. The meter is calibrated according to the calibration method and calibration constant selected in the calibration data.

If you select the hanging code or chain code calibration mode, start the belt scale after adding the simulated load. Press the start key, the meter displays.

Automatic Interval Calibration
Time Remaining XXX
XXX.X ton/hour
Grand Total XXXX ton

During the automatic interval calibration, the meter resolution is 10 times higher than normal. Wait until the remaining time is 0, the screen shows:

Automatic interval calibration completed
Error XXX.X %
Change the interval?
Change Suspend

If you need to change the interval, press the change button, the screen shows:

Interval Changed
New Interval XXXXX
Previous Interval XXXXX
Run Menu

Such as the instrument show:

— Input Error —
Max:68000000
Min: 70000
Return

Please check whether the input parameters are correct, and consult an engineer if necessary.
Operation Details

conveyor weight scale

3.1.1 LED Status Indicator
5 red LEDs indicate the current operating status of the totalizer
1.Remote indication (Flashes during auto-zero adjustment)
2.Control indication (Flashes when automatically adjusting the interval)
3.Alarm (Flashes when there is an alarm)
4.Batch process
5.Normal operation
3.1.2 Keyboard

1.Run – Enter the running state, press this key at any time, the instrument will enter the running state
2.Menu – Enter the main menu
3.Roll up/Roll down – Scrolling menu options
4.Soft key — Perform the function shown on the key
5.Number keys 0-9 – Input data parameters
6.Decimal point“.” – Input decimal point
7.C – Clear wrong input
8.Local/Remote – Select the setpoint input method
9.Auto/Manual – Setting the working mode of the regulator
10. Start – Start loading output(Effective installation of batch board)
11. End – Abort loading output(Effective installation of batch board)
3.2 Menu display
The meter uses the interactive Chinese menu to prompt the user to perform parameter setting, calibration or testing operations. In any state, you can press the menu key to cycle through the main menu 1-6, press the corresponding soft key to select the desired sub-menu and pass Scroll up/Scroll down keys for further selection.

— Main Menu 1 —
Press [Menu] to continue
Zero Interval Object
Calibration Calibration Calibration

— Main Menu 2 —
Press [Menu] to continue
Weighing Calibration
Display Data Data

— Main Menu 3 —
Press [Menu] to continue

Diagnosis

— Main Menu 4 —
Press [Menu] to continue
Input-Output Alarm
Set Set Se

— Main Menu 5
Press [Menu] to continue
Special rate Address Function

— Main Menu 6 —
Press [Menu] to continue

Control

3.3 Run display
After the instrument is powered on, it displays the running status. The running state can also be entered by pressing the run key in any menu.

3.3.1 Master Accumulation

XXXX.XX t
XXX.X t/h

Cleared

The first line displays the main accumulation, which starts to accumulate when the scale is initialized.
The second line shows the instantaneous flow rate.
The third line shows clear, press the soft key below clear, the indicator will display

Is the master total cleared?

YES NO

Press Yes to clear the main total and return to the main operation, press No to return to the running display directly.
3.3.2 Reset accumulation
Reset accumulation is used for auxiliary measurements, such as shift accumulation or batch accumulation. Press the scroll up or scroll down button to display the reset accumulation

 

XXXXX t
XXXXX t/h

Reset Cleared

Press the reset button to prompt the instrument

Is the master total cleared?

YES NO

 

Press Yes to reset the accumulation and reset to zero and return to the reset accumulation, press No to directly return to the reset accumulation display.
3.3.3 Regulator output display
Press the scroll up/down key to display the main accumulation, reset the accumulation and the regulator output in turn

XX.X t/h
XX.X t set up
■×
Confirm Conversion

The first line shows the flow
The second line shows the set point of the regulator
The third line shows the deviation between the actual flow rate and the setpoint, and each square represents a 2% control deviation.
Press the second line of the conversion screen to display in a loop
1. Set point-When the setpoint is local, it will display confirmation, you can use the number keys to enter the setpoint.
2. Error%-Display the error between the actual flow rate and the setpoint as a percentage.
3. Control output-When the regulator is in manual mode, it will display confirmation, and the control output can be manually input.
3.3.4 Alarm processing
When an alarm occurs, the right soft key displays the alarm, and the alarm LED flashes

XXXXXXX t
00.0 t/h

Alarm

Press the alarm key to display

Alarm New
XXXXXX

Reset Continue

Alarm “new” means that the alarm has not been recognized, press the reset button to reset the alarm output.
Press the Continue button to display other alarm messages that still exist.
Note: The alarm light will go out only when the fault that caused the alarm is restored.
3.4 Calibration
The main menu 1 includes the menus of “zero-point calibration”, “interval calibration” and “physical calibration”, please refer to sections 2.5.9, 2.5.10, and 4.1.5 respectively.

Chapter 4 Menu
4.1 Main Menu 1
Main menu 1 includes zero point calibration and interval calibration, press the menu key to select main menu 1

–mainmenu1 —
Press [Menu] to continue
Zero Interval
Calibration Calibration Calibration

Calibrate by zero point, the screen shows

–Automatic zero-point calibration –
Idling belt
Press the start button again
Automatic Exit Manual

4.1.1 Read current zero point and manual zero point
Press manual display

— Manual zero point —
XXXX t/h
Zero: XXXXX
Confirm Exit

Max: 120000
Minimum: 0
Default: 0

The second line displays the current flow rate, the third line displays the current zero point, use the numeric keys to enter the known correct zero point
4.1.2 See section 2.5.9 for automatic zero-point calibration
4.1.3 Reading interval and manual calibration interval

– Interval calibration –
Start belt
Press the start button again
Automatic Exit Manual

Select interval calibration in main menu 1, the screen displays
Press the manual button, the screen displays

— Manual interval adjustment —
XXXXX t/h
Interval: XXXXX
Confirm Exit

Max:68000000
Minimum: 70000
Defaults:10055293

 

The second line displays the current flow rate, and the third line displays the current zero point. Use the numeric keys to correct the interval.
4.1.4 See 2.5.10 for automatic interval adjustment
4.1.5 Physical calibration
Physical calibration means that the meter uses actual materials to calibrate

Physical calibration
The belt scale is idling and then press the start button
Start Menu

 

the belt scale. The materials must be accurately weighed on a static scale before passing through the belt scale.

Conveying a certain amount of material on the scale

Continue

In main menu 1, press physical calibration, the screen displays

XXXX.XX t
XXX.XX t/h
Press the finish button to end
Complete Abort

Start the belt scale, wait for the speed to stabilize, and press start

Press the continue key to transport the calibration materials on the belt scale, and the meter will start to accumulate at this time.

When performing physical calibration, the resolution of the cumulative sum flow is 10 times higher than normal. When all materials pass through the belt scale, press the Finish button (if you press Cancel, the physical calibration process will exit)

XXXX.X t
Enter reference weight
XXX.X t
Confirm Abort

Max: 3500.00
Minimum: 0.10

After inputting the accurate weight of the material passing through the belt scale, press the confirm key, the meter displays

Interval calibration completed
Error XXX.X%
Change the interval?
Change abort

Press the change key, the meter should display

Interval changed
New interval XXXXX
Old interval XXXXX
Run menu

 

Press the run key, the instrument enters the running state
Such as the instrument display

— input error —
Maximum value: 45000000
Minimum value: 500
return

Please check whether the input parameters are correct and consult an engineer if necessary。
4.1 Main menu 2
Main menu 2 includes the main parameter configuration menu of the system, press the menu button to display main menu 2

– Main menu 2 –
Press [Menu] to continue
Scale calibration
Display data data

4.2.1 Show
Press the display key, the screen displays

– Start scrolling – Press [Scroll] to select

4.2.2 Select the cumulative unit
Press the scroll key and the screen displays

– Display scroll 1 –
Cumulative unit
> t <
Select confirm

 

Default unit: t (ton)
Select unit: t, Kg

 

Press the select button to switch options, press the enter button to confirm the option。
4.2.3 Select flow unit
Press the volume button, the screen displays

– Display scroll 2 –
Flow unit
> t/h <
Select confirm

 

Default unit: t/h (tons/hour)
Select unit: t/h, Kg/h percentage%

Note: The percentage is the ratio of the current flow to the maximum scale capacity
4.2.4 Set display flow buffer time
The flow rate displayed on the screen can be filtered by setting the damping
time to filter out the flow jitter caused by mechanical vibration,

– Display scroll 5 –
Flow decay time
2 seconds
confirm

Press the scroll key and the screen displays

Max: 200 second
Min: 0 second
Default: 2 second

4.2.5 The third line shows
Press the scroll key and the screen displays

– Display scroll 6 –
The third line shows
> None <
Select confirm return

Speed, time, and belt volume per meter (kg/m) are optional
4.2.6 Scale data
Press the Menu key to return to the main menu 2, press the scale data key, the screen displays

4.2.7 Maximum scale capacity
Press the scroll key and the screen displays

–Scale data scroll 1 –
Maximum scale flow:
10.0 t/h
confirm

Default: 10.0
Min: 1.0
Max: 200,000.0

Use the number keys to enter the maximum scale capacity, press the enter key

4.2.8 Select scale division
Press the volume button, the screen displays

– Scale data display 2 –
Scale
> 0.1 <

Default: 1
Choose: 0.1, 0.01, 0.001, 1

Press the select key to select the desired division value, press the enter key to set

4.2.9 Select belt scale model
Press the volume button, the screen displays

– Scale data display 3 –
Scale model
> 30 <

Default: 30
Choose: 30,17,14,20-2

Press the select button to select the mode, press the confirm button to set.
4.2.10 Set zero dead zone
The zero point dead zone range is the percentage value of the maximum scale capacity. When the material flow is lower than the set zero point dead zone range, the accumulation will stop accumulating. Press the scroll key and the screen displays

– Scale data scrolling 4–
Zero dead zone
0%

Max: 5%
Min: 0%
Default: 0%

Use the number keys to input the zero point dead zone range, press the Confirm key to set。
4.2.11 Remote cumulative output index
Press the volume button, the screen displays

– Scale data scroll 5 –
Cumulative output index
0.01 tons
confirm

Max: 100
Min: 0.01
Default: 1

The remote cumulative pulse division is set according to the maximum flow when the instrument is running, and is used to control the remote counter. The meter generates count pulses in the cumulative output division on the set cumulative output port (see 4.4.3). Normally, the pulse frequency should not exceed 2.5 Hz.
4.2.12 Cumulative pulse width
Press the scroll key and the screen displays

– Scale data scroll 6-
Cumulative pulse width

Max: 20
Min: 0
Default: 0

The input value is multiplied by 0.1 seconds to obtain the actual pulse width.
4.2.13 Load cell capacity
Press the scroll key and the screen displays

– Scale data scroll 7–
Load cell capacity
150 kg
confirm

 

Max: 5000
Min: 0
Default: 150

4.2.14 Load cell sensitivity
Press the scroll key and the screen displays

– Scale data scroll 8–
Load cell sensitivity
3.0000mV/V
confirm

Max: 8
Min: 0
Default: 3

 

– Scale data scroll 9–
Number of load cells
1
confirm

Maximum value: 6
Minimum value: 1
Default value: 3

4.2.16 Define the speed signal input method
Press the scroll button to display on the screen

– Scale data display 10–
Speed input
> External <
Select confirm

Default: external
Select: external, analog

When the speed sensor is not connected, the analog speed signal function is selected. The internal analog frequency of the instrument is 20Hz speed signal for users to use.
4.2.17 Load cell resistance
Press the scroll button to display on the screen

– Scale data scroll 11–
Load cell resistance
 350Ω
confirm

Maximum value: 1000
Minimum value: 100
Default value: 350

4.2.18 Roller spacing (belt scale is invalid)
Press the scroll button to display on the screen

– Scale data scroll 12–
Roller spacing
1200.0 mm
confirm

Maximum value: 5000
Minimum value: 200
Default value: 1200

4.2.19 Max kg/m (material weight per meter)
Press the scroll button to display on the screen

– Scale data scroll 13–
Max kg/m
100kg/m
confirm

Maximum value: 2000
Minimum value: 0
Default value: 100

4.2.20 Belt scale inclination
Press the scroll button to display on the screen

Maximum value: 20
Minimum value: 0
Default value: 0

4.2.21 Calibration data
Press the Menu key to return to the main menu 2, press the calibration data key, the screen displays

– Start scrolling – Press [Scroll] to select

4.2.22 Calibration mode
Press the scroll key and the screen displays

–Calibration data scroll 1 –
Calibration mode:
》Pick-up calibration》
Select confirm

Optional: electronic calibration (not required), hanging code calibration, chain code calibration

Use the select key to select the calibration mode, press the enter key
4.2.23 Select the calibration weight
Press the volume button, the screen displays

–Calibration data scroll 6 –
Select the calibration weight
***Kilogram
Select confirm

Press the select key to select the desired value, and press the confirm key to set.
4.2.24 Hanging code calibration constant

Press the scroll button to display on the screen

– Calibration data scroll 7 –
Kilogram
confirm

Press the number keys and press the Confirm key to set.
4.2.25 See 2.5.6—2.5.8 for automatic establishment of the test cycle
4.2.26 Automatic zero tracking
In the calibration data, press the scroll key until the display is as follows:

-Calibration data scrolling 14-
Automatic zero tracking
> No <
Select confirm

There are two options: yes and no. If you select tracks, zero adjustments will be started when the flow is lower than the set value, and the tracking mark “Z” will be displayed.
Press the scroll key and the screen will display as follows:

-Calibration data scroll 15-
Zero tracking range
4.0%
confirm

Enter the tracking setting value and press the enter key.
4.3 Menu 3
Main menu 3 contains system test and diagnosis functions, press the menu button to display on the screen

– Main menu 3 –
Press [Menu] to continue

Protection diagnosis

4.3.1 diagnosis
Press the diagnostic key and the screen displays

– Diagnostic scroll 1-
XXXXX A/D
XXXX X
return

The second line shows the raw data of the weight signal through the AD converter
The third line shows the frequency (Hz/min) and the number of divisions of the speed pulse after frequency division.
4.3.2 Protection settings

Press the protection key and the screen displays

enter password
No protection
——–
Confirm back

In the “unprotected” state, the instrument parameters can be modified.
Enter the correct password (factory setting 0000) and press the confirm key, the screen displays

enter password
protect
——–
Confirm back

At this time, the instrument parameters are read-only and cannot be modified. Enter the correct password again to switch to the unprotected state.
If the password is incorrect, the screen will display

Wrong password
return

4.3.3 change the password
In the unprotected state, press the volume key, the screen displays

Change password
——–
Confirm back

Use the number keys to enter the password (within 8 digits) and press the enter key.
4.4 main menu 4
Main menu 4 is used to define digital input, digital output, and alarm functions, press the menu key to display on the screen

 

– Main menu 4 –
Press [Menu] to continue
Input-output alarm
Set set set

 

4.4.1 Input settings
Press the input setting key, the screen displays

– Digital input setting –

External alarm 1 0 NO
Confirm NC/NO continue

The mainboard of the instrument provides 3 programmable input interfaces. The screen displays the physical ports corresponding to each logic function in turn. Press the continue button to cycle through the settable logic functions. The NC/NO button selects the normal state of the input port (normally closed/normally open). ), NO means that the input port is closed and effective. Press the number keys to configure the physical port (#1, #2, #3) corresponding to the logic function, and press the enter key to complete the input. When the port number is set to 0, this function is invalid.
The following table shows the definable logic functions. The user can assign to any available input port.
Note: Only one logic function can be defined for each physical input port

Input function Default value
External alarm 1 0 NO
External alarm 2 0 NO
External alarm 3 0 NO
Belt running 0 NC
Cumulative reset 0 NC
Alarm reset 0 NC
Batch start 0 NC
Batch abort 0 NC

4.4.1Output settings
Press the output setting key in the main menu 4, the screen displays

– Digital input setting –

External alarm 1 0 NO
Confirm NC/NO continue

The mainboard of the instrument provides 3 programmable input interfaces. The screen displays the physical ports corresponding to each logic function in turn. Press the continue button to cycle through the settable logic functions. The NC/NO button selects the normal state of the input port (normally closed/normally open). ), NO means that the input port is closed and effective. Press the number keys to configure the physical port (#1, #2, #3) corresponding to the logic function, and press the enter key to complete the input. When the port number is set to 0, this function is invalid.
The following table shows the definable logic functions. The user can assign to any available input port.
Note: Only one logic function can be defined for each physical input port

Input function Default value
External alarm 1 0 NO
External alarm 2 0 NO
External alarm 3 0 NO
Belt running 0 NC
Cumulative reset 0 NC
Alarm reset 0 NC
Batch start 0 NC
Batch abort 0 NC

4.4.1Output settings
Press the output setting key in the main menu 4, the screen displays

–Digital output setting –
Alarm 0 NO
Confirm NO/NC to continue

The mainboard of the instrument provides 3 programmable output interfaces. Press the continue button and the screen will display the physical ports corresponding to each logic function in turn. The NC/NO button selects the normal state of the output port (normally closed/normally open), and NO means that the port is closed when the output is valid. . Press the number keys to input the physical port (#1, #2, #3) corresponding to the output function, and then press the enter key. When the port number is set to 0, this function is blocked.

The following table shows the definable logic functions. Users can assign to any available output port.

Note: Only one logical function can be defined for each physical output port
Output function Default value
Alarm 0 NC
Stop 0 NC
High flow 0 NC
Low flow 0 NC
*Control deviation 0 NC
Cumulative 0 NC
warn! !!
Changing the digital output settings may cause related equipment to start. The user should take care to avoid personal injury and equipment damage caused by incorrect operation.
4.4.3 Alarm setting
Press the alarm setting key in the main menu 4, the screen displays

4.4.4 High flow set point
Press the scroll key and the screen displays

Maximum value: 100
Minimum value: 0
Default value: 100

The set value is a percentage value relative to the maximum scale flow. Press the number keys to input the required set point, and press the enter key to complete the input.
4.4.5 High flow alarm delay

– Alarm scroll 2 –
High flow alarm delay
0 seconds
confirm

Press the scroll key and the screen displays

Maximum value: 50
Minimum value: 0
Default value: 0

Press the number keys to enter the delay time, and press the enter key to complete the entry.
4.4.6 Low flow alarm point
Press the scroll key and the screen displays

– Alarm scroll 3 –
The low flow set point
10%
confirm

Maximum value: 100
Minimum value: 0
Default value: 10

The set value is a percentage value relative to the maximum scale flow. Press the number keys to input the required set point, and press the enter key to complete the input.
4.4.7 Low flow alarm delay
Press the scroll key and the screen displays

Maximum value: 50
Minimum value: 0
Default value: 0

Press the number keys to enter the delay time, and press the enter key to complete the entry.
4.4.8 Control deviation setpoint
Press the scroll key and the screen displays

– Alarm scroll 5 –
Control deviation set point
10%
confirm

Maximum value: 100
Minimum value: 0
Default value: 10

The set value is a percentage value relative to the maximum scale flow. Press the number keys to input the required set point, and press the enter key to complete the input.

4.4.9 Control deviation alarm delay
Press the scroll key and the screen displays

– Alarm scroll 6 –
Control deviation alarm delay
_0 seconds
confirm

Maximum value: 50
Minimum value: 0
Defaults:

Press the number keys to enter the delay time, and press the enter key to complete the entry.

4.4.10 Alarm definition
All alarm functions of the instrument can be programmed as follows
Alarm Generate an alarm message to remind the user. If the alarm output function is set, the output can reset the alarm output until the alarm is reset or the fault is eliminated.
Downtime Generate alarm information, such as setting the stop output function to generate a non-resettable stop output signal

until the fault is eliminated.
None The alarm function is blocked
Among them, high-flow alarm and low-flow alarm programmable alarm point and delay time

– Alarm definition –
Division error
>None<
Choose confirm go on

Press the scroll key and the screen displays

Press the continue key to cycle through the alarm function, press the select key to select the desired alarm mode (alarm, shutdown, none), and press the confirm key to input.
1.Alarm function list
2.1. Division error
3.2. AD converter failure
4.3. High flow alarm
5.4. Low flow alarm
6.5. Control deviation alarm*
7.6. External alarm 1
8.7. External alarm 2
9.External alarm 3
4.4.11 Alarm reset selection
Press the scroll key and the screen displays

– Alarm reset selection –

> Automatic <
Select confirm

Automatic and manual options are available.

weigh shark belt scale

 

Convey Belt scale debugging method and operation

Step 1: Establish a test cycle: (It is not the first time to calibrate the meter or change the speed sensor, you can skip this step.)
1)To idle the belt at maximum speed, press the “Menu” key twice to display the “Menu 2” page, press the “Down ↓” key until the “Establish Test Cycle” page appears, press the corresponding soft key under “Manual”, and the page “Max Speed” appears “Start the belt” prompt, press the “Continue” button, the page displays “Enter the belt cycle length***”, press the “Enter” button, the page displays the “Enter belt rotation cycle number” prompt. Enter the number of belt weeks “*” weeks. Press the “Enter” button. Then enter the belt running time ***S, press the “confirm” button, the page will display the belt running countdown, and wait until the countdown reaches 0. The page displays “Establish Test Cycle”, press the “Continue” button. The page displays “Maximum belt speed…”, press the “Enter” button. Press the “Run” button again. The establishment of the test cycle is complete.

Step 2: Zero adjustment (tare):
1) Idling the belt at maximum speed, press the “Menu” key once, “Menu 1” appears, press “Zero Calibration”, and then press the “Start” soft key, and the instrument will count. When the timer reaches 0, the page “Change the new zero point value?” appears on the page, press the “Change” button, and then press the “Run” button. The zero point calibration is complete. Repeat the zero adjustment operation 2 times. 2) Frequent zero adjustment operation is necessary to ensure accurate measurement.

The third step, the first method: hanging code calibration:
After the zero adjustments are completed, stop the conveyor belt, hang up the 10KG code (depending on the specific weight), start the belt at the maximum speed, press the menu button once, “Menu 1” appears, press “Interval Calibration”, then press the “Start” button, the instrument Proceed to time. When the timer reaches 0, the page “Change the new interval value?” appears on the page, press the “Change” button, and then press the Run button. The interval check is complete. Repeat the interval calibration operation twice. Record the interval value at this time and press the “Run” key to exit。

The third step and the second method: physical verification:
First accurately weigh out the physical weight T, after the zero adjustments are completed, perform physical verification, in menu 1, press physical verification, and when the red mark passes the mark point, press start. At this time, the material will start to pass, and wait until all the materials pass the belt scale. , And wait until the red mark passes the mark point, press the finish button, enter the standard weight T, and press the change button to complete the calibration. Go through the material again according to the above method and observe the error between the displayed weight and the actual weight.
Note: In the third step, one can be selected reasonably according to the use environment of the field equipment.

The key formula of belt scale: In menu two-interval calibration, press Manual, there will be an interval value. This interval value can be changed manually. Specific formula: new interval value=(actual weight/display weight)*old interval value or new interval value=(1±error to be adjusted)*old interval value.
For example, increase the amount displayed on the instrument by 0.5%, and the new interval value=1.005*old interval value. For example, reduce the amount displayed on the instrument by 0.5%, and the new interval value=0.995*old interval value.